Diagram
6
EC Electrode
Probes process signal directly for noise free measurements
Auto sensor recognition
Store calibration speci c data from the last calibration
Are built with materials suitable for use in chemical analysis
Have integrated temperature measurement
Incorporate a 3 mm jack termination
Unique serial ID in every probe for traceability
Probe
Probe
Diagram
Speci cations 35
EC TDS Salinity
Range
0.00 to 29.99 μS/cm,
30.0 to 299.9 μS/cm,
300 to 2999 μS/cm,
3.00 to 29.99 mS/cm,
30.0 to 200.0 mS/cm,
up to 500.0 mS/cm,
absolute
conductivity ‡
0.00 to 14.99 ppm (mg/L),
15.0 to 149.9 ppm (mg/L),
150. to 1499. ppm (mg/L),
1.50 to 14.99 g/L,
15.0 to 100.0 g/L,
up to 400.0 g/L
absolute TDS (with
0.80 factor)
0.0 to 400.0% NaCl *,
2.00 to 42.00 PSU *,
0.0 to 80.0 g/L *
Resolution
0.01 μS/cm,
00.1 μS/cm,
1 μS/cm,
0.01 mS/cm,
00.1 mS/cm
0.01 ppm,
0.1 ppm,
1 ppm,
0.01 g/L,
0.1 g/L
0.1% NaCl,
0.01 PSU,
0.01 g/L
Accuracy
@ 25 °C / 77 °F
±1% of reading
(±0.05 μS/cm or 1
digit, whichever is
greater)
±1% of reading
(±0.03 ppm or 1 digit,
whichever is greater)
±1% of reading
Calibration
Single Cell Factor calibration; 6 standards
available: 84, 1413 μS/cm, 5.00, 12.88, 80.0,
111.8 mS/cm, 1 point offset: 0.00 μS/cm
Single point with
HI 7037L Standard
Temperature
compensation
Automatic -5 to 100º C (23 to 212º F)
NoTC - none
absolute conductivity.
Conductivity
temperature
coefficient
0.00 to 6.00% / ºC (for EC and TDS only). Default value is 1.90% / ºC
TDS factor
0.40 to 0.80 (default value is 0.50)
EC Probe
HI 763100
Log feature
Up to 1000 * (400 basic mode) records organized in: Manual log on demand
(Max. 200 logs), Manual log on stability (Max. 200 logs), Interval logging *
(Max. 600 samples; 100 lots)
Temperature
Range
-20.0 to 120.0 ºC ; -4.0 to 248.0 °F**
Temperature
Resolution
0.1 ºC; 0.1 ºF
Temperature
Accuracy
±0.5 ºC; ±0.9 ºF
Speci cations
* Standard Mode Only
** Temperature limits will be reduced to actual probe limits.
‡ Absolute conductivity (or TDS) is the conductivity (or TDS) value without temperature
compensation.