Probe Calibration
DSO-2150 USB Calibration
1. When manufactured, DSO-2150 USB is calibrated manually to obtain
maximum performance.
2. You may calibrate DSO-2150 USB 5 or 6 months after purchase.
Probe Compensation
1. You must use a probe with more than 60MHz bandwidth to get undistorted
signal.
2. Probe should be compensated whenever it is connected for the first time.
3. Connect calibration signal to channel 1, then push AUTOSET.
4. Check the shape of the displayed waveform.
5. Adjust the probe until the displayed waveform is compensated.