Appendix A Specifications
All technical specifications are suitable for HDG3000B(C) series signal generators.Unless
otherwise stated, all technical specifications are guaranteed upon the following two conditions.
The signal generator is in the calibration period.
The signal generator has operated continuously for more than 30 minutes at the specified
operating temperature (18℃ to 28℃).
All specifications used are guaranteed, except those marked "Typical".
Memory Depth 2M
Maximum
100MHz 80MHz 60MHz 40MHz 25MHz 15MHz
Voltage
16Bit
Standard
Waveforms
Sine, Square, Ramp, Pulse, Noise, Harmonic, DC
Arbitrary
Waveforms
More than 160 kinds, including Sinc, Exponential Rise, Exponential Fall,
ECG, Gauss, Haver Sine, Lorentz, Dual-Tone, etc.
Frequency Characteristics
Sine
1uHz~100MHz
Square
1uHz~15MHz
1uHz~15MHz
1uHz~2MHz
Harmonic
1uHz~50MHz 1uHz~40MHz 1uHz~30MHz 1uHz~20MHz
Noise (-3dB) 100 MHZ bandwidth
Arbitrary
1uHz~20MHz 1uHz~20MHz 1uHz~20MHz 1uHz~15MHz
Accuracy
±1ppm, 18 ~ 28 ℃
Rise/Fall time
Typical (1kHz, 1Vpp)
Overshoot
Typical (100kHz, 1Vpp)
≤5%
Duty Cycle
Range varies with frequency
Non-symmetry 1% of the period + 4ns
User Manual 82