6
360094-1/J
QuickThaw™ Plasma Thawing System Operation Manual
3 Installation
3.1 Location requirements
► A level surface
► Clear of temperature extremes
► Minimum of 8 inches (203 mm) above
► Access to grounded outlet
► Access to water supply
► Meets limits specied for ambient temperature and relative humidity
► Adjacent drain or waste container
NOTE: If possible, drain or waste container should be on same side as drain port. Otherwise, U-tube and
180 degree tting will extend drain tubing to other side.
3.2 Power on
1 With thawer at room temperature, connect power cord to thawer and to proper grounded outlet.
2 Press power button on control panel (ON/OFF toggle).
For DH8 100 V models, the POWER button also controls the fan.
NOTE: High temperature alarm activates quickly when chamber is empty.
3.3 Install external thermometer
Optional. Install per manufacturer documentation. If using a Helmer DT1 digital thermometer, see
instructions on the CD.
3.4 Specications
Input voltage and frequency
The requirements for a particular thawer are specied on the Product Specication label. The voltage
tolerance is ±10% of the nominal voltage.
Options:
► 100 V, 50 Hz
► 115 V, 50 Hz to 60 Hz
► 230 V, 50 Hz to 60 Hz
Power consumption
Power consumption is measure in full load Amperes.
Input voltage
Model
DH2 DH4 DH8
100 V
3.0 A 8.0 A 12.0 A
115 V
2.5 A 6.0 A 10.0 A
230 V
1.25 A 3.0 A 5.0 A