22
Table 2. Service
Parameter Manual test
mode
Autom. Test
mode
Survey of
errors
Adjust Putting into
operation
Update
Operating hours P-scheduled Auto-Test Survey of errors Basic position Serial number Update
Counter total V-scheduled Casting position Manufact. date
Counter since Pressure PDC:W Software version
Min. pressurisation Chamber UDC 1:V Checksums
Max. pressurisation Cooling UDC 2:V Operating hours
Max. oscill. temp. IDC:A Counter total
8.5.1 Description Table 2. Service
Parameters Inthemenuparameterssomeunit-specicvaluesareincludedthatarerelevantfortheuser
or service technician, e.g. the total operating hours, a counter for the total number of castings
and a counter that registers castings from a certain date (e.g. last service) on.
The pressurisation times and the max. oscillator temperature are particularly useful for service
technicians since they allow to conclude on connection conditions of the laboratory and
possible faulty castings. These values can only be reset by service technicians.
Manual test mode Only service technicians have access to the manual test mode. All individual steps of the
Auto-Test which is run when the unit is switched on can be tested manually.
Automatic test mode This menu has only one selectable function “Auto-Test”.
The Auto-Test can be started or interrupted with this key.
Error history This item provides the user with a survey of the last 15 error messages displayed by the casting
machine. The adjustment knob is used to scroll through the messages.
Adjustment of the chamber This option is only available for service technicians. The upper and lower stop of the chamber
can be corrected electronically. Incorrect adjustment results in error messages and might even
damage the mechanic system for turning the chamber!
Putting into operation Serial number and manufacturing date have been stored in this menu.
They can only be entered or changed by the manufacturer. The software version and the
checksums depend on the current software version used in your machine and can not be edited.
If changes or manipulations performed by the customer should be determined, any liability
claim and guarantee shall lapse.