Calibration, Controls and Daily Maintenance
Calibration
Ensure that [Warming Up], [Lamp Warming Up], or [Ready to Test] is displayed.
• For new slide lot #, insert the QC card into the slot with the QR code facing to the left.
NOTE: Test name and lot #’s appear on the display and will print out. Analyzer will store 5 lots per analyte or panel.
Controls
NOTE: Always wear gloves when running controls.
1. Run HESKA
Chemistry Control solution as you would a patient sample by transferring into a 0.5-ml Red Top tube.
2. Select [Control] for your species.
Daily Maintenance
1. Keep cover closed except when loading and unloading slides, samples and consumables.
2. Empty waste disposal drawer.
3. Change reference fluid every 12 hours.
4. Remove sample tubes when finished testing.
5. Clean any spills on the spotting slide transit area.
qs
quick steps guide
Element DC5X
Veterinary Chemistry Analyzer