Element DC5X
Veterinary Chemistry Analyzer
Element DC5X™ Veterinary Chemistry Analyzer Supplies
NOTE: For access to the full user manual for this product, please visit www.heska.com/productmanuals.
Remove testing supplies from refrigerator and warm to room temperature:
Slides: 5 minutes in foil pouch.
NOTE: Remove only those needed for ordered tests.
Reconstituted HESKA® Chemistry Control: 5 minutes.
Electrolyte Reference Fluid: 5 minutes.
NOTE: Fill 0.5-ml or 1.5-ml Red Top sample tube depending on the reference fluid setting designated in, Tube
setting for diluent & reference fluid” (Administrator mode), then return remaining RE Fluid to refrigerator.
Replacing Consumables
1. First ensure [Warming Up], [Lamp Warming Up], or [Ready to Test] is displayed on the screen.
2. To stop sampling, touch STOP.
Setting Tips
1. Touch the image of the tip racks on the display screen to turn sample carousel.
2. Once carousel stops, remove tip racks.
3. Fill tip racks and replace.
4. Touch RESET, then touch FINISH.