HTS-420R Maintenance Manual
9
List of Common Maintenance Items
Long level vial calibration
circular level vial calibration
Telescope reticle calibration
Collimation error calibration(2C calibration)
Index error calibration(2I calibration)
Compensator mechanical zero position alignment
Compensating linearity calibration of compensator
Prism/No Prism mode switch motor
light path switch motor replacement
Maintenance Operations
1. Long vial calibration
1) Fix the instrument on a solid and stable platform and set the vial to the position between leveling screw
A and B, then continue the steps below.
2) Adjust the screw A and B to center the bubble. Turn the instrument horizontally by about 90°and then
adjust screw C to center the bubble. Turn the instrument 180°to check if the bubble is still staying in the
center. If it is, the vial is aligned, otherwise take the next step.
3) Use the adjusting pin to twist the screw beside the vial to make the bubble move half distance toward
the center.
4) Repeat the step 2 and 3 until the bubble is in the center in every direction.