During startup inspection, the operator intentionally applies electric current to check
whether the 3159 and the test leads are functioning properly.
(1)
(2)
(3)
(4)
Connect the power cord (supplied with the
3159) to the inlet of the main unit and a
grounded outlet.
Always properly ground the 3159 in order
to prevent electric shocks.
Short-circuit the high-voltage test lead
(red) and low-voltage test lead (black) as
shown in the picture.
Wear high-voltage rubber gloves when connecting the lead wires.
Connect the test leads securely so that they do not disconnect during the test,
Do not place the test leads directly on the ground or floor during the test. Do not
bring them into contact with any metal products in the vicinity.
Place the connected test leads on a board
made of refractory insulated material, such as
an electric-proof rubber sheet.
Check that the power switch of the 3159 is
turned off and the output voltage knob is in the
zero position.
Tightly attach the supplied low-voltage test
lead (black) to the LOW terminal (black).
Insert the supplied high-voltage test lead (red)
to the HIGH terminal (red) all the way so that it
is firmly seated.
2. Simple Startup Inspection (Connection)