8
(1) When the result is “PASS”
(2) When the result is “FAIL”
(3) If the FAIL Hold function is enabled, press the STOP (EXIT) key to cancel the hold state.
<After the test>
(1)
(2)
(3) Turn the main power switch of the 3159 off.
(4) Disconnect the electrical appliance, test leads, and power cord in the reverse order of
connection.
An "UPPER FAIL" judgment will be given and the 3159 will beep if the electric
a
liance fails to withstand the test volta
e for the set test time.
Record the result as FAIL in the record form or other documents.
(If the measured voltage value, leakage current value, and remaining time are also
necessar
record them as well.
Return the output voltage knob to zero. (Turn the knob
all the wa
counte
-clockwise
Check that the analog voltmeter indicates
0 kV
and that
the
lamp is
not illuminated
.
To conduct further tests, start from “5. Preparation Before the Test” of this measurement guide.
A "PASS" judgment will be given if the electric appliance withstands the test voltage for the set
test time. (“PASS” will be displayed for approximately 0.5 seconds by default; however the
“PASS” indication can be retained. See “4.1 PASS Hold Function” in the 3159 instruction
manual.)
Record the result as PASS in the record form or other documents.
7. Finish the Test
Measured voltage value
Remaining time
Leakage current value
Measured voltage value Remaining timeLeakage current value