Hioki 3159 - Safety Notes

Hioki 3159
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
(1)
(2)
Set the test voltage after pressing the START key to start the test.
Check the values to be used before
p
ressin
g
the START ke
y
.
When a test is started, high voltage is generated by the 3159 when it enters the state.
Do not touch the output voltage HIGH terminal (red), test leads, electrical appliance (device to
be tested), or short-circuited receptacle while the lamp is illuminated.
The test requires the voltage to be applied for 60 seconds.
After pressing the START key, turn the output voltage knob to set the voltage
to the test voltage before the remaining time reaches 60 seconds.
To Set the Test Voltage
To set the test voltage, turn the output voltage knob while checking the
measured voltage value.
It may take time to get used to setting the test voltage.
Turning the output voltage knob excessively can raise the
voltage over the test voltage, resulting in damage to the
electrical appliance.
Press the START key to start the test.
To forcibly terminate the test process,
press the STOP (EXIT) key.
6. Start the Test
Measured voltage value Leakage current value Remaining time