2
Setting screen
Set reference value
(example 10.5Ω)
Change
sign:
Return to
Measurement
screen
Set upper tolerance
(example +4.5%)
Set lower tolerance
(example -4.5%)
Decide According to
Reference Value and Tolerance
When setting the upper tolerance, the
lower tolerance is automatically set to
its negative equivalent. Change the
setting as needed.
Select item
Cancel:
Setting Comparator Decision Criteria for Measured Values
REF% Mode
12
4
3
Change
numerical value:
Return to
Measurement
screen
Decide According to
Upper/Lower Thresholds
Set upper thresholds
(example 150mΩ)
Set lower thresholds
(example 50mΩ)
Setting screen
Select item
Change
numerical value:
Cancel:
ABS Mode
12
4
3
+OvrRng
-OvrRng
Contact Check Fault
Indicates possible contact fault or broken
cable.
Voltage Level Monitor Fault
Probe contact may be unstable, or electri-
cal noise ingress may be present.
No Measurement Data
No measurement value is present imme-
diately after power on or after changing
range. Errors with other causes can pre-
vent display of measured values.
Measurement Fault (Error) Indicators
Measurement Screen Indicators
Trigger Source
INT: Internal trigger
EXT: External trigger
Measurement
range
Measurement speed
Appears only when OVC is enabled
(OVC: Offset Voltage Compensation)
Appears only when the Low-Power
Resistance function is enabled
Key-Lock State
F.LOCK: Changes to settings are disabled
M.LOCK: All except comparator settings are disabled
Zero-adjust (ON)
Remote control
C.E. Hi
C.E. Lo
C.E. Volt
- - - - - - -
Cancel key-lock
(hold 1s)
PRINT
Prints measured value
(Enabled when PRINT is selected as the INTERFACE
setting on the SYSTEM screen)
PRINT
Prints statistical data
(Enabled when PRINT is selected as the INTER-
FACE setting on the SYSTEM screen)
STAT
Displays statistical calculation results
(Enabled when ON is selected as the STATISTICS set-
ting on the DATA screen)
UNDO
Deletes the previously stored measurement
and calculation result (only the most recent
measurement can be deleted)
NUMBER
Sets the number of measurements for Auto-Memo-
ry mode
(Enabled when ON is selected as the AUTO MEMORY
setting on the DATA screen)
ALLCLR Deletes all memories and calculations
Enable local
control
Tenkey Input Enabled
Change Settings
Out of Range (over- or under-range)
The measured value is above or below
the measurement range, or open device.