-6-
MN1-5531 Rev. 10
Caution
Attach the waterproof cover if the connector will be soaked in liquid.
If the connector is soaked in liquid without the waterproof cover, liquid can get inside and result in
malfunction.
Attach the waterproof cover to the probe and use it by following the instructions in the waterproof
cover manual. Using the wrong combination of the probe and the waterproof cover, incorrectly
mounting the waterproof cover, or using the probe without properly mounting the cover can cause
the probe to malfunction.
Do not use a waterproof cover where a problem has been found.
Using a waterproof cover in an abnormal state can cause injury to the user. Contact one of our ofces
and/or distributor's ofces listed on the back cover.
Do not use the waterproof cover in sterilization that applies changes in pressure.
Using ethylene oxide gas sterilization, low-temperature plasma
sterilization, or other sterilization methods that apply changes in pressure can
damage the probe and the waterproof cover. In these cases, do not use the waterproof cover.
After soaking in chemical solution, check that no liquid has entered into the connector.
If liquid appears to have entered the connector, immediately stop use and contact one of our ofces
listed on the back cover.
Do not use the waterproof cover if the packing has been removed before.
The waterproof cover will not function correctly even if packing that was removed is returned to its
original location. Replace it by a new waterproof cover.
When cleaning the waterproof cover, do not deform the packing by applying unnecessary force.
Use of the waterproof cover while the packing is deformed can cause liquid to enter the probe
resulting in malfunction.
Do not rub the packing of the waterproof cover with a brush.
This could damage the packing. Use of the waterproof cover while the packing is damaged can cause
liquid to enter the probe resulting in malfunction.
Be sure to store the waterproof cover by removing it from the connector.
If the waterproof cover is stored while connected to the probe, the packing can become deformed.
Replace the waterproof cover around two years after it is commercing for use.
The packing for preventing liquid deteriorates with time.
For safe use, replace the waterproof cover by a new one typically two years.