-37-
MN1-5531 Rev. 10
5-7. Automated cleaning and disinfecting
Probe (with waterproof cover)
The following items must be provided prior to automated cleaning and disinfection:
a) Washer disinfector: according to ISO 15883 with chemo-thermal program (temperature: max 60 °C [140 °F]).
b) Waterproof cover for probe connector MP-2790
c) Detergent: Korsolex Endo-Cleaner (Bode Chemie; # 972 020)
d) Disinfectant: Korsolex Endo-Disinfectant (Bode Chemie; # 972 030)
1) The parameters of the cleaning and disinfection of the device are as follows:
Program step
Water
(40l)
Dosage
(ml/l)
Temp.
(°C)/(°F)
time
(min)
Pre-Rinse Cold water 5
Cleaning Deionized water 5 (0.5%) 50/122 10
Rinse Deionized water 1
Disinfection Deionized water 10 (1%) 55/131 5
Rinse Deionized water 1
Rinse Deionized water 55/131 1
Drying 55/131 15
2) Connect the waterproof cover MP-2790 to the probe connector and conrm there is no air leak by the
tester. About detail information, refer to the instruction manual of the waterproof cover MP-2790.
3) After closing the door, start the chemo-thermal program
4) Open the door after the end of the program.
5) Check whether the probe is dry. If not, proceed as described under drying The applicable chemical
solutions are listed below.