EasyManua.ls Logo

Hoffer Flow Controls HIT-4U - Inputs; DC Power;Loop Powered; Analog Output

Hoffer Flow Controls HIT-4U
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Features and Specifications 9
HIT-4U HP-327
2.2 Inputs
Magnetic Pickup:
Frequency Range: 0.2 Hz to 5000 Hz.
Signal Level: 30 mV
P-P
to 30 V
P-P
.
Opto-Isolated DC Pulse:
Frequency Range: 0 Hz to 3000 Hz.
Signal Level: 0 to +DC pulse.
Internal Pull-Up 10 k to +DC
Low (Logic 0): < 1 VDC
Min Pulse width: 0.1 msec
Contact Closure:
Frequency Range: 0 Hz to 5000 Hz
Internal Pull-up: 220 k to +3.3 VDC
Reset:
Signal Type: Contact closure
Min Time On: 25 msec
Internal Pull-up: 35 k to +3.3 VDC
External Magnet
MCP/RF Pickup
2.3 DC Power/Loop Powered
Voltage: 8 to 30 VDC
Current: < 24 mA
Loop Burden: 8 VDC maximum
Supply Backup: Two (C) Size Lithium or battery pack for
Ex systems
Battery Life: 2 years typical-Two (C) size battery
1 year typical-Ex system battery pack
Protection: Reverse polarity protected
2.4 Analog Output
Scale: 4 – 20 mA follows rate.
Accuracy: 0.02% of Full Scale @ 20ºC.
Temperature drift: 40 ppm/°C
Update Time: 0.125 seconds.
Protection: Reverse polarity protected
Flow-Tech, Inc. | In MD 410-666-3200 - In VA 804-752-3450 | www.flowtechonline.com