48 HIT-4 Communication Program
HP-327 HIT-4U
Table Points:
Set the number of points to be used for the linearization table.
Valid entries are 2 to 20.
Average K-Factor:
Enter the average flowmeter K-Factor. Valid entries are 0.001
to 9999999.9.
Correction Factor:
Enter flow and total multiplier. Valid entries are 0.001 to
99999.999.
K-Factor Method:
Select flowmeter linearization method as “Average” (single
K-Factor) or “Linear” (2 to 20 point linearization table), or
"Viscosity".
K-Factor Units:
The K-Factor Units selection provides a way to enter calibration
data in various units of measure. The units of measure must be
selected prior to entering the K-Factors. When K-Factors are
written to the device, they are converted to and stored in the base
units pulses/gallon. The K-Factor Units parameter is provided
for convenience and is not stored in the device.
Outputs Page
The Outputs Configuration screen is used to configure the Analog, Alarm
and Pulse outputs.