Honeywell Analytics CM4 Protocol Manual
Get the Number This command allows you to query the unit for the
of Gas Tables number of loaded gas tables in the unit.
Available 0x3A
Command packet to instrument:
Command Code (0x3A) - 1 byte
Response packet:
Command Code (0x3A) - 1 byte
CM4 Date - 2 bytes
CM4 Time - 2 bytes
# of Gas Tables - 1 byte (1-255)
Get Printer This command queries the unit for the printer
Setup conguration.
0x3B
Command packet to instrument:
Command Code (0x3B) - 1 byte
Response packet from instrument:
Command Code (0x3B) - 1 byte
CM4 Date - 2 bytes
CM4 Time - 2 bytes
Setup Status - 1 byte
bit 0: Printer port enable/disable
0 = Disable
1 = Enable
bits 1-2: Printer report format
00 = Continuous
01 = Summary
10 = Compressed
11 = Invalid
bits 3-5: Printer baud rate
000 = 1200
001 = 2400
010 = 4800
011 = 9600
100 = 19200
bit 6: Printer hardware handshaking
(ow control)
0 = Disabled
1 = Enabled
bit 7: Undened