35 Rev. 2.4 (3/05)
Honeywell Analytics CM4 Protocol Manual
Set Printer This command congures the printer for output.
Conguration
0x57
Command packet to instrument:
Command Code (0x57) - 1 byte
Setup Status - 1 byte
bit 0: Printer port enable/disable
bits 1-2: Printer Report format
00 = Continuous (prints all conc. for
each pt)
01 = Summary (prints alarms, faults,
and TWA only)
10 = Compressed (prints conc. At or
above LDL)
11 = Invalid
bits 3-5: Printer baud rate
000 = 1200
001 = 2400
010 = 4800
011 = 9600
100 = 19200
bit 6: printer hardware handshaking (ow
control)
0 = Disabled
1 = Enabled
bit 7: Undened
Response packet from instrument:
Command Code (0x57) - 1 byte
CM4 Date - 2 bytes
CM4 Time - 2 bytes
Status Flag - 1 byte
0x00 = Printer programmed
0x01 = Invalid report format
0xff = Printer programming error