38 Rev. 2.4 (3/05)
Honeywell Analytics CM4 Protocol Manual
Set TWA Time This command congures the TWA time output.
0x5A You need to enter only the initial TWA output
time. The other two time factors are calculated
automatically.
Command packet to instrument:
Command Code (0x5A) - 1 byte
TWA Time - 2 bytes
Command Code (0x5A) - 1 byte
CM4 Date -2 bytes
CM4 Time - 2 bytes
Status - 1 byte
0x00 = Value saved and veried
0x01 = Hours Invalid
0x02 = Minutes Invalid
0xff = Not saved
Set Display This command congures the length in seconds
Cycle Time that the concentration for each point is displayed
0x5B while in the Monitor mode.
Command packet to instrument:
Command Code (0x5B) - 1 byte
Cycle time - 1 byte (valid 2-10 seconds)
Response packet from instrument:
Command Code (0x5B) - 1 byte
CM4 Date - 2 bytes
CM4 Time - 2 bytes
Status - 1 byte
0x00 = Value saved and veried
0x01 = Value < 2 seconds
0x02 = Value > 10 seconds
0xff = Not saved