EasyManuals Logo

Honeywell Enraf VITO User Manual

Honeywell Enraf VITO
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
Average temperature measurement with MPT / RTD
___________________________________________________________________________
Instruction manual VITO Page 34
4.4.3 Temperature pointer (items VP and VV)
By means of the value pointer (item VP) a vector can be loaded to the HCU or ICU_HPI option board. Next, with
item VV, the selected data is returned. Item VP consists of 4 positions, in the middle separated by a ‘.’ or ,’ :
v w . x y (or v w , x y). The values for the value pointer are listed in table 4 – 2 together with the obtained data.
v w
,
x y
Selected data
Example / Dimension
0 0
0 0
0 0
0 0
0 2
0 2
:
0 2
0 2
0 2
0 2
0 2
:
0 2
0 3
0 3
0 3
0 3
0 3
,
,
,
,
,
,
,
,
,
,
,
,
,
,
,
,
,
0 0
0 1
0 2
0 3
0 0
0 1
:
1 3
2 6
2 9
8 2
8 3
:
9 5
0 1
0 9
1 0
1 1
1 2
HCU / ICU_HPI Emulation & Function
Emulation: MPU, HSU, HPU, HCU
Function: HC: HART channel installed
AO: analog output
ST: spot temperature
MR: VITO average MRT/multi spot probe
WS: external water bottom probe
PR: pressure transmitters
Configuration boot code
Sales code option: C boot code: 0A
J 2E
W 1A
X 12
Y 3F
HCU / ICU_HPI hardware version
Boot code software version
Calculated temperature of 1
st
spot element (floating point format)
Calculated temperature of 2
nd
spot element (floating point format)
:
Calculated temperature of 14
th
spot element (floating point format)
Standard deviation (floating point format)
Calculated I_ref (floating point format)
Calculated resistance 1
st
spot element (floating point format)
Calculated resistance 2
nd
spot element (floating point format)
:
Calculated resistance 14
th
spot element (floating point format)
Error counters HART addresses 3, 4 and 5
(762 VITO MRT Interface has HART address 5)
Detected HART device addresses (5 for VITO Interface)
Initial detected VITO device type & software version
Description: 763 : 762 VITO MRT/multiple RTD interface
RTD PRB : Multiple RTD’s
RTD : 2 or 3 RTD’s
3-W : 3-wire configuration
10 : software version 1.0
Real active functions of VITO Interface (refer to description of VP=03,10)
Actual executed HART commands from 762 VITO Interface
VV=HCU HCAOMR- - PR
VV=HCU CONFIG: 3F
VV=HW VERSION: 00
VV=BOOTSW VERS:01
according to item TD
according to item TD
according to item TD
< 0.5E-10
1.0275 mA ±0.05
VV=0000:0000:0013
VV= - - - - 5 - - - - - - - - -
VV=763 RTD 3-W 10
VV=763 RTD 3-WIRE
Table 4 – 2
Item Name Description
VP= Temperature value pointer Temperature value pointer; format: 2 digits, separator, 2 digits
(refer to table 4 – 2).
Example: VP=02.00: value pointer loaded to request the
calculated temperature of the 1
st
spot element.
VV Temperature pointer value Temperature pointer value.
This item holds the value requested by item VP
(refer to table 4 – 2).
Example: the requested temperature of the 1
st
spot element
(standard floating point format).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Enraf VITO and is the answer not in the manual?

Honeywell Enraf VITO Specifications

General IconGeneral
BrandHoneywell
ModelEnraf VITO
CategoryTemperature Controller
LanguageEnglish

Related product manuals