EasyManua.ls Logo

Honeywell STT700 STT700 - User Manual

Honeywell STT700 STT700
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Honeywell Process Solutions
STT700 SmartLine
Temperature Transmitter
User’s Manual
34-TT-25-17
Revision 1
July 2017

Table of Contents

Questions and Answers

  • S
    Steven WardAug 4, 2025
    Why Honeywell Transmitter has no input or low input?
    • K
      Kathryn GarzaAug 4, 2025
      If you're experiencing no input or low input on your Honeywell Transmitter, the issue could stem from a faulty sensor wire connection or an incorrect operating voltage. First, carefully inspect the field wiring and connections. Then, make sure the transmitter is operating within its correct voltage range. Lastly, check that the 250 ohms resistor is properly installed in the loop.
  • M
    mark83Aug 18, 2025
    What to do if Honeywell Transmitter shows high input?
    • M
      Mr. Scott OconnellAug 18, 2025
      If your Honeywell Transmitter shows high input, it might be due to a bad sensor wire connection. Inspect the field wiring and connections for any partial short circuits. Also, verify that the transmitter operates within its specified voltage range.

Summary

About This Manual

Release Information

Manual release details and publication number.

References

Support and Contact Information

Details for obtaining technical support and contacting Honeywell.

Symbol Descriptions and Definitions

Symbol Descriptions

Definitions and explanations of symbols used throughout the document.

1. Introduction

1.2. Features and Options

Lists the available models, protocols, HMI, materials, approvals, and mounting options.

1.3. STT700 SmartLine Transmitter Nameplate

Details the information found on the transmitter's nameplate.

1.4. Safety Certification Information

Information on hazardous area approvals listed on the nameplate.

2. Application Design

2.2. Safety

Covers safety aspects, including accuracy and diagnostic messages.

2.2.2. Diagnostic Messages

Lists and describes critical and non-critical diagnostic messages for the STT700.

3. Installation and Startup

3.3. Mounting

Details various mounting options for the STT700 module.

3.4. Wiring a Transmitter

Provides guidance on connecting the STT700 transmitter for operation.

3.4.6. Lightning Protector

Information on the lightning protection device for surge protection.

4. Startup

4.1.3. Output Check Procedures

Details procedures for checking the transmitter's analog output.

4.1.4. Constant Current Source Mode Procedure

Step-by-step guide for performing a current loop test.

5. Operation

5.2. Configuration Tools

Covers configuration tools like SFC and HART communicators.

6. Maintenance

6.3. Troubleshooting

Provides guidance on troubleshooting issues with the STT700 transmitter.

6.3.2. Troubleshooting with HART Communicator

Lists conditions and recommended actions for troubleshooting using a HART communicator.

7. Calibration

7.2. Calibration Procedures

Outlines calibration procedures for analog mode and refers to other manuals for HART/DE.

Appendix A. Product Certifications

A2. European Directive Information (EU)

Provides the EU Declaration of Conformity for the STT700.

A5. Marking ATEX Directive

Details information provided as part of the ATEX directive labeling.

A.6 WARNINGS and Cautions

Lists important warnings and cautions for equipment use.

A.6 Conditions of Use for Ex Equipment, Hazardous Location Equipment or Schedule of Limitations

Specifies conditions for using Ex equipment and limitations for hazardous locations.

Honeywell STT700 STT700 Specifications

General IconGeneral
TypeSmart Temperature Transmitter
Accuracy±0.1% of span
HousingAluminum or Stainless Steel
Ambient Temperature Limits-40 to 85°C (-40 to 185°F)
Hazardous Area ApprovalsATEX, IECEx, FM, CSA
InputRTD, Thermocouple, mV, Ohm
Output4-20 mA HART
Power Supply10.5-42 V DC
Stability0.1% of URL per year
Response Time<1 second
HART ProtocolHART 7
Process Connection1/2" NPT, M20 x 1.5
Electrical Connection1/2" NPT, M20 x 1.5
Wetted Materials316L SST, Hastelloy C
Relative Humidity0-100% RH

Related product manuals