What to do if Honeywell ST 800 shows sensor over temperature?
- ZZachary LeeSep 23, 2025
Verify the environment temperature is within specification. Take steps to insulate Meter Body from temperature source.

What to do if Honeywell ST 800 shows sensor over temperature?
Verify the environment temperature is within specification. Take steps to insulate Meter Body from temperature source.
Why is my Honeywell Transmitter showing PV out of range?
If the Honeywell Transmitter displays 'PV Out of Range', it might be due to a sensor overload or fault, a redundant characterization calculation error, or the calculated pressure exceeding the Upper Transducer Limit (UTL). Check the range and, if necessary, replace the transmitter with one that has a wider range. The meter body may be damaged, so check the transmitter for accuracy and linearity. Replace the Meter Body and recalibrate if needed.
How to fix a Honeywell ST 800 Transmitter with a display failure?
If the Honeywell Transmitter's display is not working, ensure the display connections are secure. If the issue continues, try resetting the device. If the problem persists after the reset, consider replacing the display.
What does loop current noise mean on a Honeywell Transmitter?
If you observe loop current noise frequently on your Honeywell Transmitter, it could indicate an impending critical under or over-current failure. Closely monitor the device for other failure signs, or proactively replace the electronics module.
What causes a sensor fault in Honeywell ST 800?
If the Honeywell Transmitter shows 'Sensor Fault', it could be due to a failure on the sensor board, a redundant characterization calculation error, or the calculated pressure exceeding the Upper Transducer Limit (UTL). Check the range and, if required, replace the transmitter with one that has a wider range. The meter body may have been damaged. Check the transmitter for accuracy and linearity. Replace the Meter Body and recalibrate if needed.
What to do if Honeywell ST 800 Transmitter shows URV set error?
Verify the inputs are valid for the intended operation.
What does excess zero correction mean on Honeywell ST 800 Transmitter?
If the Honeywell Transmitter displays 'Excess Zero Correction', check the input pressure to ensure it matches the calibrated range values (LRV and URV). Check the Meter Body. Perform an LRV or ZERO CORRECT procedure followed by a URV CORRECT procedure.
How to fix PILD blockage detected on Honeywell Transmitter?
To resolve a 'PILD Blockage Detected' error on your Honeywell Transmitter, inspect and clear the impulse lines. Then, disable and re-enable PILD when the process and transmitter are operating normally.
What to do if Honeywell ST 800 reports no DAC compensation?
If the Honeywell Transmitter displays 'No DAC Compensation', replace the Electronics Module (PWA) to achieve the maximum current loop accuracy or return the device to factory for DAC compensation. Note that loop accuracy may be slightly compromised, resulting in a minor degradation of ambient temperature influence specifications.
What does 'No Factory Calibration' mean on a Honeywell ST 800 Transmitter?
If the Honeywell Transmitter is showing 'No Factory Calibration', it means the factory calibration is missing, which will compromise accuracy. Replace the meter body or return the device for factory calibration.