Revision 5.0 ST 800 Series HART/DE Option User’s Manual Page 79
SV Tracking
Table 17Maximum SV Tracking
Max SV Limit Parameter Description
Meterbody Temperature (SV) upper
operating limit from specification.
Set-up
None.
Max SV
Value
Parameter
Description
Highest Meterbody Temperature ever
experienced by the device.
Set-up
None - value initialized to Min SV Limit
value prior to leaving the factory.
Updates to current Meterbody
Temperature automatically when
powered at user site after one minute.
NVM
Update after every 8 hours.
Time Above
Upper Stress
Limit
Parameter
Description
Accumulation of minutes that device’s
Meterbody Temperature has been
above the value of “SV Upper Stress
Limit”.
Set-up
None.
NVM
Backup once each 8 hour period
Time Since
Last SV Up
Method
Description
Time that has passed since the last
time device’s Meterbody Temperature
has passed above the value of “SV
Upper Stress Limit” (in days, hours and
minutes).
Set-up
None.
NVM
Backup once each 8 hour period
Table 18Minimum SV Tracking
Min SV Limit
Parameter
Description
Meterbody Temperature (SV) lower
operating limit from specification.
Set-up
None.
Min SV
Value
Parameter
Description
Lowest Meterbody Temperature ever
experienced by the device.
Set-up
None - value initialized to Max SV Limit
value prior to leaving the factory.
Updates to current Meterbody
Temperature automatically when
powered at user site after one minute.
NVM
Update after every 8 hours.
Time Below
Lower Stress
Limit
Parameter
Description
Accumulation of minutes that device’s
Meterbody Temperature has been
below the value of “SV Lower Stress
Limit”.
Set-up
None.
NVM
Backup once each 8 hour period
Time Since
Last SV
Down
Method Description
Time that has passed since the last time
device’s Meterbody Temperature has
passed below the value of “SV Lower
Stress Limit” (in days, hours, and
minutes).
Set-up
None.
NVM
Backup once each 8 hour period

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell ST 800 Series and is the answer not in the manual?

Honeywell ST 800 Series Specifications

General IconGeneral
BrandHoneywell
ModelST 800 Series
CategoryTransmitter
LanguageEnglish

Related product manuals