Touchpoint 4 Serial Communication Protocol Issue 1 09/07
Reset fault & alarm – 0x41
This command allows a remote reset of latched visual, audio and relay outputs resulted
from an alarm or fault.
Command packet to instrument:
Commandcode: 0x41
Data field: null
Example:
Ifthemasterwantstoresetlatchedoutputsofaddressed01instrument,itshould
send out:
0x7f0x010x010x410x3e
Response packet from instrument:
Commandcode: 0x41
Data field: null
Example:
Theaddressed01instrumentreceivestheresetfault&alarmcommand
and responses:
0x7f0x010x010x410x3e