SINGLE-GAS DETECTOR
50115903-101 EN B2
QUICK REFERENCE GUIDE
BW Clip Real Time is a portable, single-gas detector manufactured by Honeywell. BW
Clip Real Time continuously displays the concentration of a specific gas in the ambient
environment, and activates an alarm when concentrations exceed alarm setpoints. It is
your responsibility to respond appropriately to the alarms.
BW Clip Real Time is a member of the GasAlert family of products. It is compatible
with IntelliDoX automatic test and calibration stations, as well as Safety Suite
Device Configurator software. For more information, visit the product website at
www.honeywellanalytics.com
Safety Information: Read First
a WARNING
1. Substitution of components may impair Intrinsic Safety.
2. Honeywell recommends performing a bump test prior to each day’s use to confirm
sensor response and alarm activation by exposing the detector to a concentration
of target gas that exceeds the low alarm set point. Honeywell also recommends
performing a bump test if the detector has been subjected to physical impact, liquid
immersion, an Over Limit alarm event, custody changes, or anytime the detector’s
performance is in doubt. Calibrate the detector after a failed bump test. Always bump
test and calibrate in a fresh air environment.
3. Detectors in hibernation mode are not functional.
a CAUTION
1. Activate the detector before the activation date on the package.
2. This product is a gas detector, not a measurement device.
3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
4. Clean the exterior with a soft, damp cloth.
5. For optimal performance, periodically zero the sensor in a normal atmosphere
(20.9% v/v O
2
) that is free of hazardous gas.
6. Portable safety gas detectors are life safety devices. Accuracy of ambient gas
reading(s) is dependent upon factors such as accuracy of the calibration gas
standard used for calibration and frequency of calibration. Honeywell recommends
performing a calibration at least once every 180 days (6 months).
Informations de sécurité – À lire au préalable
AVERTISSEMENT
1. Le remplacement d’un composant de l’appareil peut compromettre la sécurité
intrinsèque du détecteur.
2. Avant chaque utilisation quotidienne, Honeywell recommande d’effectuer un test
fonctionnel afin de vérifier la réaction du capteur et l’activation de l’alarme, en
exposant le détecteur à une concentration de gaz cible supérieure au seuil d’alarme
basse. Honeywell recommande également d’effectuer un test fonctionnel si le
détecteur a été soumis à un impact physique, à une immersion dans du liquide, à un
événement d’alarme Dépassement de la limite, à des changements d’utilisateur ou
chaque fois que les performances du détecteur sont mises en doute. Étalonnez le
détecteur après l’échec d’un test fonctionnel. Effectuez toujours un test fonctionnel
et étalonnez dans un environnement avec de l’air frais.
3. Détecteurs en mode veille prolongée ne sont pas fonctionnels.
BW Clip Factory Calibration Certificate
This instrument was factory inspected, tested and calibrated in accordance with the conditions
and requirements of our registered Quality System, Operating Standards and Sales Agreements.
MISE EN GARDE
1. Activez le détecteur avant la date d’activation sur l’emballage.
2. Ce produit est un détecteur de gaz, pas un dispositif de mesure.
3. Veillez à assurer que la grille du capteur est exempt de saleté, les débris, et ne soit
pas vobstrué.
4. Nettoyez l’extérieur de l’appareil à l’aide d’un chiffon doux et humide.
5. Pour des performances optimales, zéro périodiquement le capteur dans une
atmosphère normale (20,9% v/v O2) qui est exempte de gaz dangereux.
6. Les détecteurs portables de gaz de sécurité sont des dispositifs de sécurité des
personnes. La précision des valeurs de gaz ambiant dépend de divers facteurs, tels
que la précision de la norme de gaz d’étalonnage utilisée pour l’étalonnage et la
fréquence d’étalonnage. Honeywell recommande d’effectuer un étalonnage au moins
une fois tous les 180 jours (6 mois).
BW Clip Real Time Factory Calibration Certificate
This instrument was factory inspected, tested and calibrated in accordance with the conditions
and requirements of our registered Quality System, Operating Standards and Sales Agreements.
Getting Started
ButtonButton
Sensor and
sensor grill
Sensor and
sensor grill
Bubble label
with product name
and gas type
Bubble label
with product name
and gas type
Audible
alarm
Audible
alarm
LCDLCD
LED
visual alarms
LED
visual alarms
Infrared port
Alligator
clip
Product label
LCD and Symbols
Button prompt. When this symbol is displayed, press and hold the
button until the symbol disappears.
BW Clip Real Time O
2
models only. Oxygen concentration is
measured as percent by volume.
BW Clip Real Time H
2
S, CO and SO
2
models only. Toxic gas
concentration is measured in parts per million.
Low alarm and high alarm. These symbols are displayed when a gas
concentration that exceeds alarm setpoints is detected.
Non-compliance warning. This symbol is displayed when a
diagnostic or compliance test fails; when the operating life
countdown reaches 24 hours or less; and for up to 30 days after
operating life expiry. It is also displayed when certain notifications
are enabled through Safety Suite Device Configurator via an
IntelliDoX station.
Real-time clock. This symbol is displayed when real-time clock is
enabled through Safety Suite Device Configurator via an IntelliDoX
station.
End of operating life warning. This symbol is displayed when the
operating life countdown reaches 24 hours or less.
Peak reading. This symbol is displayed when a maximum gas
exposure event has occurred within the past 24 hours.
Bump test reminder. If the detector is configured to display a bump
test reminder, this symbol is displayed when a bump test is due.
indicates a failed bump test.
This symbol is displayed when more than 24 hours have elapsed
since the last successful zero procedure was performed (O
2
models) or indicates a sensor calibration is due (H
2
S, CO and SO
2
models).
indicates a failed calibration (H
2
S, CO and SO
2
models) or a
failed zero procedure (O
2
models).
Operating life remaining in months (MM) or days (DD).
Question and Answer IconNeed help?

Do you have a question about the Honeywell BW Clip Real Time and is the answer not in the manual?

Honeywell BW Clip Real Time Specifications

General IconGeneral
BrandHoneywell
ModelBW Clip Real Time
CategoryGas Detectors
LanguageEnglish

Summary

Safety Information

General Safety Warnings and Cautions

Important safety precautions and warnings for detector use and maintenance.

Getting Started

Detector Components and Overview

Identifies key components, displays, and indicators of the BW Clip Real Time detector.

Understanding Detector Symbols

Explains the meaning of various symbols displayed on the detector's LCD screen.

BW Clip Factory Calibration Certificate

Factory Calibration Details

Details regarding the instrument's factory inspection, testing, and calibration.

Detector Activation and Operation

Activating and Operating the Detector

Step-by-step guide to activate the detector and its normal operating status.

Understanding Operating Life Indicators

Explains the countdown display for remaining operating life and expiry warnings.

Diagnostic and Alarm Functions

Internal Diagnostic Tests and Fail-Safe

Describes internal diagnostic tests and the fail-safe procedure if a test fails.

Alarms and Warnings

Details low, high, over-limit, and under-limit alarms and their indicators.

Non-compliance Symbol and LED Indicators

Explains the non-compliance warning symbol and associated LED flashing patterns.

Detector Settings and Events

Automatic Zero Reminder and Peak Readings

Covers automatic zero reminders for O2 models and display of peak gas readings.

Recalling Events and Settings

How to scroll through stored events and detector settings using the button.

Detector Procedures

Zeroing the Sensor

Procedure for zeroing the sensor, especially for O2 models, and troubleshooting failed zero.

Bump Tests

Guidelines for performing bump tests, including IntelliDoX station usage.

Calibration

Instructions for calibrating the detector with recommended gas mixtures and troubleshooting.

Event Logs

Details on stored events, their content, and how to transfer them.

Optional Settings and Hibernation

Optional Settings Configuration

Enabling optional settings like non-compliance alerts and real-time clock display.

Activating Hibernation Mode

Instructions for activating hibernation via a case or Safety Suite Device Configurator.

Specifications and Certifications

Contact and Warranty Information

Contact details, warranty registration, and limited warranty information.

Detector Specifications

Details on maximum operating life, operating temperatures, detection ranges, and alarm setpoints.

Ratings and Certifications

Lists the standards and certifications the BW Clip Real Time detector complies with.

Recommended Calibration Gas Mixtures

Specifies recommended gas mixtures for H2S, CO, O2, and SO2 calibration.