07/2019 - 4007779101 - 34-VF-25-76 iss.1 GLO July 2019 US 2
2 Electrical connection
Minimum inlet section
15 DN
50 DN
20 DN
20 DN
40 DN
30 DN
> 5 DN
5 DN
5 DN
Minimum outlet section
Heat insulation for T
medium
> +160°C / +320°F
1
1
2
3
1 Marking of max. height of the insulation
2 Max. thickness of the insulation up to the bend of the pressure pipe
3 Insulation
1
2
3
4
5
6
7
8
1 Open the housing cover of the electrical terminal compartment using the key
2 Signal converter supply and 4...20 mA loop
3 4...20 mA current input, - external transmitter, optional
4 Terminal M1 binary (high current)
5 Terminal M3 binary (NAMUR)
6 Terminal M2/4 binary, common minus connection
7 Ground terminal in housing
8 Ground terminal on connection piece between flow sensor and signal converter
Connecting the signal converter
Sunshade