34
8.14 Hi-pot test "Arc Sensitivity" Setting Interface
In this interface, press the "Input" key. The first digit of the voltage value will be
displayed in reverse. Use the numeric keys to set the required arc sensitivity. After
setting, press “Enter” to confirm. To abandon the setting, press the "Esc" key to exit
the current setting.
Arc sensitivity setting options:
0: Turn off the arc alarm function
1-9: Turn on the arc alarm function, the higher the number, the higher the sensitivity
9. Continuous test function setting (AC and DC Hi-pot test and insulation
resistance test)
Use the "▲" or "▼" key on the measurement parameter setting interface to
select the "continuous test" setting content, as shown in the figure below:
8.15 Hi-pot test "Connect" setting interface
Connect setting options:
NO: Disable the continuous test function of this group
YES PASS: Enable the continuous test function of this group and connect to the next
group of tests when the test passes.
YES ALL: Enable the continuous test function of the group, no matter the test results
is pass or fail and connected to the next group of tests.
The continuous function of this group is enabled, that is, the connection to the next
group is completed when the conditions of this group of tests are satisfied.
For example, to test the connection of M1, M2 and M3, you only need to enable the