What to do if Hospira Lifecare PCA3 Medical Equipment displays 'Pendant Fault - O'?
A
armstrongchristineSep 12, 2025
If the Hospira Medical Equipment displays 'Pendant Fault - O', it could be due to several reasons:
* The pendant is not securely inserted into the port.
* There is a faulty connection.
* The pendant is only partially pressed.
* The incorrect pendant is being used (PCA Plus II Patient Pendant - white handle).
To resolve:
* Secure the pendant into the port.
* Replace the pendant or pump.
* Release the pendant button.
* Confirm that you are using the PCA 3 Patient Pendant (blue handle).
L
Lucas RobertsSep 13, 2025
What to do if Hospira Lifecare PCA3 Medical Equipment displays 'Pendant Fault - S'?
A
Angela MooreSep 13, 2025
If the Hospira Medical Equipment displays 'Pendant Fault - S', it indicates an electrical short in the pendant. To resolve this, replace the pendant.