EasyManua.ls Logo

HP 120A - Horizontal Amplifier; Calibrator; AC or DC Coupling

HP 120A
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sect,
n
Page
3
Select
the
two
centimeter
portion
to
be
magnified
with
the
HORIZONTAL
POSITION
control.
Trigger
Level
-
Normally,
a
trace
will
be
displayed
in
the
absence
of
a
signal,
since
the
oscilloscope
is
triggered
automatically.
However,
you
may
select
the
point
of
triggering
and
stop
the
AUTOMATIC
triggering
with
the
TRIGGER
LEVEL
control.
Move
this
front
panel
screwdriver
control
clockwise
away
from
the
AUTOMATIC
switch
position
(extreme
coun
terclockwise)
only
for
special
triggering.
In
the
AU
TOMATIC
position
the
sweep
is
self-triggered
in
the
absence
of
a
triggering
signal.
When
a
signal
is
fed
in,
the
signal
controls
the
triggering
action.
In
the
AUTOMATIC
position
the
trigger
circuit
is
adjusted
to
optimum
setting
to
trigger
on
the
majority
of
wave
forms
and
at
a
zero
trigger
level.
As
you
.rotate
the
control
clockwise
from
the
AUTOMATIC
switch
po
sition
the
automatic
sweep
feature
will
stop.
Further
rotation
erf
this
control
permits
you
to
adjust
the
trig
gering
level
of
the
pattern,
within
limits.
2-6
HORIZONTAL
AMPLIFIER
The
SWEEP
TIME
-
HOR.
SENS,
switch
also
con
-
trols
the
sensitivity
of
the
horizontal
amplifier
when
it
is
used
with
an
external
input.
The
three
sensiti
vity
ranges
of
the
horizontal
amplifier
are
.
1,
1,
and
10
VOLTS/CM.
The
sensitivity
of
the
amplifier
may
be
increased
to
100
VOLTS/CM
by
means
of
the
VER
NIER
control.
Even
though
the
circuitry
of
the
vertical
and
horizon
tal
amplifiers
is
not
identical,
the
bandshapes
and
phase
characteristics
are
identical,
at
least
to
well
above
100
kc.
As
a
result
the
instrument
is
rated
as
having
not
more
than
differential
phase
shifts
be
low
100
kc
and
is
thus
well
suited
to
phase
difference
measurements.
In
fact,
because
the
horizontal
ampli
fier
has
basically
an
even
wider
response
than
the
vertical,
an
adjustment
(C107,
Phase
Adjustment)
is
provided
in
the
horizontal
amplifier
so
that
phase
differences
can,
if
desired,
be
minimized
at
indivi
dual
frequencies
above
100
kc.
The
adjustment
con
sists
of
a
small
variable
capacitor
which
acts
to
set
the
first
high-frequency
corner
of
the
horizontal
am
plifier
for
optimum
agreement
with
that
of
the
verti
cal.
At
the
factory
this
adjustment
is
set
for
opti
mum
agreement
at
100
kc.
2-7
CALIBRATOR
An
internal
square-wave
calibrator
with
an
accuracy
of
±2%
and
a
nominal
frequency
of
400
cps
is
provided
for
checking
amplifier
calibration
accuracy.
The
cal
ibrator
is
connected
internally
to
the
input
of
the
ver
tical
amplifier
when
the
VERT.
SENSITIVITY
switch
is
placed
in
the
CAL.
position.
When
in
the
CAL.
po
sition
a
60
millivolt
square
wave
is
fed
into
the
ver
tical
amplifier
input
for
calibration
purposes.
Be
cause
of
the
inherent
stability
and
medium
sensitivi
ty
of
the
horizontal
amplifier,
no
provision
is
made
for
internal
connection
between
that
amplifier
and
the
calibrator.
Horizontal
amplifier
calibration
can
be
checked
easily,
however,
since
an
external
signal
can
be
measured
on
the
vertical
amplifier
and
then
applied
to
the
horizontal
as
a
check.
Little
adjustment
of
the
horizontal
amplifier
is
necessary.
In
either
automatic
or
trigger
operation
the
sweep
circuits
wiU
trigger
from
signals
over
a
range
from
600
kc
to
below
5
cycles
(50
cps
for
automatic
opera
tion).
Between
10
cps
(50
cps
for
automatic)
and
200
kc
the
minimum
deflection
needed
for
triggering
from
displayed
waveforms
is
0.
5
cm
peak-to-peak.
Be
yond
this
range
an
increase
in
this
minimum
will
occur,
but
at
600
kc,
for
example,
triggering
will
usually
occur
on
waveforms
that
give
2
cm
deflec
tion.
For
external
trigger
signals
a
minimum
am
plitude
of
2-1/2
volts
peak-to-peak
is
required
for
the
minimum
frequency
to
200
kc
ranges.
2-8
EXTERNAL
SYNCHRONIZATION
CONTROLS
Switch
the
SYNCHRONIZATION
control
to
the
EX
TERNAL
position.
Feed
in
the
external
synchro
nization
signal
(2.
5
volts
or
more
peak-to-peak)
into
the
horizontal
input
terminals
at
the
lower
right
corner
of
the
front
panel.
The
external
synchro
nization
signal
is
always
ac
coupled
and
synchro
nizes
on
the
negative
slope
of
the
signal.
When
using
the
trigger
level
control
the
trace
will
trigger
on
the
negative
slope
of
signals
within
the
range
of
+10
and
-10
volts,
approximately
(see
Figure
2-2).
Figure
2-2.
Triggering
2-9
AC
OR
DC
COUPLING
AC
coupling
permits
high
gain
to
be
employed
with

Related product manuals