Example You want to load a configuration file called “486_bus” from the hard disk of
the HP 16500C into an HP 16550A state/timing module. The HP 16550A is
installed in slot B of the HP 16500C mainframe. To load the configuration file
from the HP 16500C hard disk, you need to send the programming command
to the analyzer. The syntax of the command is:
:MMEMory:LOAD:CONFig <filename>[,<disk drive>][,<slot number>]
In this case, the disk drive parameter, will be “INT0,” which indicates the
hard disk. The slot number will be 2, because the HP 16550A is installed in
slot B. To load the configuration file, enter the following command at the
DOS prompt:
C:\> echo :MEM:LOAD:CON ’486_BUS’,INT0,2 L:\system\program
If you are using a UNIX system, you can use the UNIX echo command.
See Also The HP 16500C System Programmer’s Guide for more information on
programming command syntax.
To copy files to the setup.raw file locations, or to send commands to the system
at the \system\program file location, you must be connected to the HP 16500C
system as the control user.
File Management
To set up the HP 16500C
3–15

Table of Contents

Other manuals for HP 16500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 16500C and is the answer not in the manual?

HP 16500C Specifications

General IconGeneral
BrandHP
Model16500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals