Calibration
3457A
5-9.
In
the
event
that
the
security
code
is
unknown,
the
security
feature
can
be
disabled
to
permit
a
new
code
to
be
entered.
To
disable
the
security
feature,
complete
the
following
steps.
a.
Remove
all
power
from
the
3457A
(line
cord
and
external
inputs)
and
turn
the
unit
upside-down.
b.
Loosen
the
six
captive
screws
in
the
bottom
cover.
(It
is
not
necessary
to
remove
these
screws
from
the
bot-
tom
cover).
c.
Return
the
unit
to
an
upright
position
and
remove
the
top
cover.
d.
Remove
the
NORM
/
DSA
/
KEY
jumper
(JM532)
from
the
NORM
position
and
place
it
in
the
KEY
position.
(JM532
is
located
on
the
AI
logic
board
directly
behind
the
front
panel
display
).
e.
Reconnect
the
power
and
turn
the
instrument
ON.
f.
Access
the
SECURE
command.
(Press
the
Blue
SHIFT
key
then
the
RECALL
/
S
configuration
key,
Use
the
|
scroll
key
to
display
the
SECURE
command
).
g.
Enter
the
number
0,
the
delimiter
(,
)
and
the
security
number
you
wish
to
use.
h,
Press
the
ENT
key.
i,
Disconnect
power
and
return
jumper
JM532
to
the
NORM
position.
j.
Replace
the
top
cover,
tighten
the
retaining
screws
and
reconnect
the
power.
The
instrument
will
now
respond
to
the
security
code
just
entered,
When
jumper
JM532
is
in
the
KEY
position,
the
security
feature
is
disabled.
It
is
possible
to
calibrate
the
instrument
without
entering
a
security
number
under
these
conditions,
If
a
new
security
number
is
not
entered
while
the
jumper
is
in
the
KEY
position,
the
original
number
will
again
be
in
effect
when
jumper
JM
532
is
return-
ed
to
the
NORM
position.
5-10,
Calibration
Number
(CALNUM?)
5-11.
The
CALNUM?
command
provides
you
with
a
method
of
monitoring
the
number
of
calibrations
per-
formed.
By
using
this
command,
you
can
determine
whether
un-authorized
calibrations
have
occurred.
The
instrument
will
always
respond
the
CALNUM
query
command
by
displaying
the
number
of
calibration
RAM
entries
completed
since
factory
initialization.
factory
initialization.
The
maximum
calibration
num-
ber
which
can
be
stored
is
32767.
After
reaching
the
maximum
number,
the
counter
resets
to
0
and
begins
again,
Please
note
that
the
calibration
number
is
increased
by
one
for
each
calibration
point.
A
complete
calibration
increases
the
calibration
number
by
several
digits.
5-12,
The
procedure
for
reading
the
Calibration
Number
is
as
follows:
a.
Access
the
CALNUM?
command.
(Press
the
Blue
SHIFT
key
then
the
OFFSET
COMP
/
C
configuration
key.
Use
the
|
scroll
key
to
display
the
CALNUM?
command).
b.
Press
the
ENT
key.
5-2