Satisfactory
completion
of
the
self
test
routine
is
indicated
by
a
display
message
of
SELF
TEST
OK.
If
an
error
is
detected,
the
display
message
will
be
TEST
FAILED
and
the
front
panel
ERR
annunciator
will
flash.
This
indicates
the
ERR?
register
contains
a
nonzero
value.
If
the
ERR?
command
returns
a
value
of
one
(I)
there
is
a
hardware
error
(see
ERR?
command
in
Command
Reference
Section
of
HP
3457A
Operating
Manual).
Another
method
of
determining
the
failure
mode
is
to
read
the
ERROR
register
(Presy
the
Blue
SHIVT
key
then
the
ERROR
key).
\f
there
is
a
hardware
error,
examine
the
AUXERR
register
by
executing
the
command
AUXERR?.
A
number
will
be
displayed
that
indicates
the
specific
test(s)
that
failed.
This
number
corresponds
to
self
test
failure
information
in
Table
8-2.
If,
for
example,
the
number
32
is
displayed
on
the
control
panel
when
the
AUXERR?
command
is
ex-
ecuted,
Table
8-2
shows
that
AUXERR
32
corresponds
to
a
Current
Source,
Gain
Selection,
Input
Divider
failure.
If
more
than
one
test
fails,
the
AUXERR
numbers
will
be
summed
together.
For
ex-
ample;
if
the
Front
end
zero
measurement
test
and
the
AC
attenuator
flatness
test
failed,
the
AUNERR
number
displayed
would
be
272
(256
+
16).
For
troubleshooting
purposes,
the
lowest
numbered
AUXERR
failures
should
be
repaired
first.
Left
of
the
AUXERR#,
in
Table
8-2,
is a
TEST#.
This
TEST#
corresponds
to
the
test
descriptions
provided
in
the
Self
Test
(TEST)
paragraphs
in
this
section,
Table
8-2.
Self
Test
Failure
Information
TEST#
AUXERR#
FAILURE
DESCRIPTION
N/A
1
Isolation
error
in
normal
operation
(not
during
self
test)
1
[
2
A/D
controller
self
test
error
2
4
Isolation
self
test
error
3
8
Integrator
convergence
error
4
16
Front
end
zero
offset
error
5
32
Current
source,
gain
select
or
input
divider
error
ai
|
6
64
Amps
function
error
4
128
AC
amplifier
DC
offset
error
8
256
AC
attenuator
flatness
error
9
512
Ohms
precharge
circuit
error
0
1024
U502
ROM
checksum
error
0
2048 U503
ROM
checksum
error
0
4096
Nonvolatile
RAM
(U511)
failure
0
8192
Volatile
RAM
(U506)
failure
0
16384
Calibration
Ram
(U511)
or
RAM
protection
circuit
error
*