PERFORMANCE
TEST
CARD
90
DAY
LIMITS
Hewlett-Packard
Model
3457A
Digital
Mutimeter
Serial
Number
DC
CURRENT
TEST
Test
Performed
by
Date
Reference
Temperature.
Step#
Input
to
Set-Up
and
High
Reading
Low
Test Test
34570
Configuration
Limit
Limit
Pass
Fail
OFFSET
TEST
1
Open
300
UADCI
Range
+
000.0104
uA
=
000.0104
uA
2.
Open
3
mADCI
Range
=
+
0.000104
ma
+
0.000106
me
3
Open
30
mADCI
Range
+
00.00104
ma
=
00.00104
me
4
Open
300
mADCI
Range
+
000.0204
mA
000.0204
ma
5
Open
1
ADCI
Range
+
0.000604
A
-
0.000604
A
ey
ene
GAIN
TEST
6
300
uA
300
uADCI
Range
300.0704
uA
299.9296
UA
7
3ma
3
mADCI
Range
3.000704
mA
2.999296
ma
8
30
mA
30
mADCI
Range
30.00704
ma
29.99296
mA
9
300
ma
300
mADCI
Range
300.2304
mA
299.7696
ma
10
1A
1
ADCI
Range
1.001304
A
0.998696
A
a
ae
AC
CURRENT
TEST
Step#
Input
to
Set-Up
and
High
Reading
Low
Test
Test
34578
Configuration
Limit
Limit
Pass
Fail
GAIN
TEST
1-30
ma
30
mAACI
Range
30.10300
mA
29.89700
me
2
300
mA
300
mAACI
Range
301.0300
mA
298.9700
mA
1A
1
AACI
Range
1.005300
A
0.994700
A
==