i
T
.O
. 33K8-4-551-1
4
.2
.8 Connect the TI Input HI and LO terminals to the C2 and CI leads on any open Rx channel of the Resistance
Measurement System.
4
.2
.9 Connect the TI fl Sense HI and LO terminals to the P2 and P1 leads on the same open Rx channel of the
Resistance Measurement System.
4
.2
.10 If the DMM Switch is available, set the DMM Switch to the DMM (source) position.
4
.2
.11 Select the Low Thermal Quad Scanner (P/O 2
.7) channel switch corresponding to the open Rx channel
connected to the TI in steps 4
.2 .8 and 4 .2
.9.
4
.2
.12 Select the Low Thermal Quad Scanner (P/O 2
.7) channel switch corresponding to the first Standard Resistor
value of the Applied column of Table 5.
4
.2
.13 The TI display must indicate within the corresponding values listed in the Limits column of Table 5.
4
.2
.14 Manually select the next TI range value listed in the Range column of Table 5, if applicable.
4
.2
.15 Repeat steps 4
.2
.11 through 4
.2
.14 for each remaining value listed in the Range and Applied columns of
Table 5.
Table S.
Range (S2) Applied (SI)
Limits (ppm of rdg)
10
10
AFPSL Cert Value ±20
.15
100 100
AFPSL Cert Value ±17
.19
1 k 1 k
AFPSL Cert Value ±10
.73
10 k 10 k
AFPSL Cert Value ±10
.72
100 k 100 k
AFPSL Cert Value ±10
.78
1 M *
1 M
AFPSL Cert Value ±18
.16
10 M 10 M
AFPSL Cert Value ±61
.42
* Remove the TI 1 s delay setting and press Offset Comp (S2) switch to OFF.
4
.2
.16 If the DMM Switch is available, set the DMM Switch to the Bridge (measure) position.
4
.2
.17 Disconnect the test setup.
4
.2
.18 Connect the TI Input and fl Sense HI terminals to the A terminal and the TI Input and Q Sense LO terminals to
the B terminal of the Resistance Standard (2
.11)
.
NOTE
Grounding/Shielding/Guarding variations between the TI and Resistance Standard
may be required to obtain a stable and accurate TI indication.
1
4
.2
.19 Set the Resistance Standard (2
.11) switches for 100 MS2.
4
.2
.20 Manually select the first TI range listed in the Range column of Table 6.
22