4
.3
.10 The TI display must indicate within the corresponding values listed in the Limits column of Table 7.
4
.3
.11 Set the Calibrator OPR/STBY switch to STBY.
4
.3
.12 Select the next Resistance Standard (2
.10) as listed in the Resistance Standard column of Table 7.
4
.3
.13 Repeat steps 4
.3
.5 through 4
.3
.12 for each remaining value listed in Table 7 using the appropriate Resistance
Standard (2
.10)
.
Table
7.
Range (ADC)
Resistance Standard
Applied (ADC)
Limits (ADC)
100
1
.1
1 kS2/Step (10
k12)
100 p
.
99
.99707 to 100
.00293
µ
1 m
100 SI/Step (1 k12)
1 m
0
.9999737 to 1
.0000263 m
10m
1 IcQ/Step (100
S2)
*
10 m
9
.999737 to 10
.000263 m
100 m
100 1
-
2/Step (10 S2) *
100 m
99
.99593 to 100
.00407 m
1
10 S2/Step (112) *' **
1
0
.9998797 to 1
.0001203
* Connect 10 Resistors in parallel.
** Heat build-up in the 1 S2 shunt [Standard Resistor (2
.10)] will affect accuracy
. If test exceeds one minute,
press Calibrator OPR/STBY switch to STBY and wait five minutes before repeating test.
4
.3
.14 Disconnect the test setup.