T
.O
. 33K8-4-551-1
4 .4
.31 The TI display must indicate within the corresponding values listed in the Limits column of Table 10.
4
.4
.32 Repeat steps 4
.4
.30 and 4
.4
.31 for each remaining applied value listed in the Applied column of Table 10.
Table 10.
Range
(AC)
Applied (MHz) Limits (VAC)
1 V
3 0
.959300 to 1
.040700
7 0
.959200 to 1
.040800
10
0
.849000 to 1
.151000
4
.4
.33 Set the Calibrator OPR/STBY switch to STBY.
4
.4
.34 Disconnect the test setup.
4
.5 AC VOLTAGE CALIBRATION
: (Analog Mode)
4 .5
.1 Connect equipment as shown in Figure 3.
4 .5
.2 Set the TI to ACV, SETACV
ANA mode, AC Band at 10 Hz to 2 MHz and resolution at
.01 by pressing the TI
S
ACV
ke s in the following sequence
:
(S HI FT)
(Blue)
11
,
press -
v
until AC BAND is displayed, then press
Enter
(SHIFT)
(Blue)
T
1
Ehter
R
E
062
SHFI
B1P)
0
Enter
(PERDD)
ccMin
I '
4 .5
.3 Set the TI to the first range listed in the Range column of Table 11.
4
.5
.4 Set the AC Measurement Standard to measure the first value listed in the Applied column of Table 11 from
INPUT 2.
4 .5
.5 Set the Calibrator for the first value listed in the Applied column of Table 11 and set the OPR/STBY switch to
OPR.
4
.5
.6 Edit the Calibrator controls until the AC Measurement Standard indicates the value listed in the Applied
column
of
Table 11.
4 .5
.7 The TI display must indicate within the corresponding values listed in the Limits column of Table 11.
4 .5
.8 Set the Calibrator OPR/STBY switch to STBY.
4
.5
.9 Repeat steps 4
.5
.3 through 4
.5
.8 for each remaining range and applied value listed in Table 11.
NOTE
Grounding/Shielding variations may be required to obtain a stable and accurate
TI indication
. The TI LFILTER may be used in Table 11 for frequencies up
through 50 kHz . Disengage TI LFILTER for frequencies above 50 kHz
. To
engage TI LFILTER
press the TI keys in the following sequence:
ENTER
I
(SHIFT)
(Blue)
L
29