Vertical
Position
Softkey
Position
Softkey
The
position
softkey
is
the
fourth
softkey.
Selecting
position
assigns
position
as
the
active
function.
You
can
use
the
position
knob
to
change
the
position
setting
at
any
time,
even
if
you
are
in
a
different
menu
or
when
position
is
not
the
active
function.
When
position
is
the
active
function,
you
can
use
either
the
position
knob
or
the
entry
knob
to
change
the
position
setting
of
the
selected
channel.
position
moves
the
trace
vertically
up
or
down
similar
to
the
vertical
position
adjustment
on
an
analog
oscilloscope.
However,
because
digital
oscilloscopes
have
a
true
dc
offset
at
the
front
end,
they
provide
a
much
wider
offset
range.
When
the
Show
key
is
pressed,
the
position
voltage
is
shown
at
the
right
of
the
waveform
display.
The
position
value
is
the
voltage
level
at
the
vertical
center
of
the
screen.

Other manuals for HP 54520 Series