CAUTION
See
Also
Vertical
Coupling
Softkey
Coupling
Softkey
The
coupling
softkey
has
several
selections:
dc,
dc
BW
lim
(dc
bandwidth
limited),
ac,
ac
BW
lim
(ac
bandwidth
limited),
and
ac
LF
reject
(ac
low-frequency
reject).
Because
the
50
Q
selection
sinks
more
current
than
the
1MQ
setting,
make
sure
you
do
not
exceed
the
maximum
rated
input
of
the
channel
when
switching
from
1MQ
to
50
Q.
When
dc
is
selected,
1
MQ
and
50
Q
dec
input
impedances
are
available
as
choices
for
input
impedance.
When
ac
is
selected,
only
1
MQ
is
available
for
input
impedance.
Bandwidth
limit
is
switchable
for
both
ac
and
dc
coupling.
Bandwidth
limit,
LF
reject,
and
ac
coupling
filter
both
the
vertical
and
trigger
paths.
e
dc
The
de
bandwidth
of
the
scope,
which
is
500
MHz.
e
dc
bandwidth
limit
Adds
a
30
MHz
low-pass
filter.
e
ac
Adds
a
10
Hz
high-pass
filter.
e
ac
bandwidth
limit
Band-pass
filter
that
consists
of
a
10
Hz
high-pass
filter
and
a
30
MHz
low-pass
filter.
e
LF
reject
Adds
a
400
Hz,
high-pass
filter.
"Edge
Trigger
Mode"
in
chapter
5
for
information
about
ac
coupling
or
adding
LF
reject
to
the
trigger
path
separately
from
the
vertical
path.

Other manuals for HP 54520 Series

Related product manuals