Operating
Errors
2000
{
2999
2019
Illegal marker type
Cause: The command executed does not apply to the marker mode of the markers on the
display.For example, if a normal marker is on screen, and the \Marker delta into span"
function is activated, the function cannot be executed, and the error will occur.
Cure:
Refer
to
the
programming
or
operation
manual
for
the
function being executed to
determine
the
proper
marker
type
for
the operation.
2020
No active marker
Cause: This error occurs when Marker Delta is moved to Center Frequency Step Size (MKSS)
while no markers are active.
2021
Bad IF/ENDIF nesting
Cure: Use the DEBUG command to locate the programming error. Refer to the
HP 70000
Modular Spectrum Analyzer Programming Manual
.
2022
REPEAT/UNTIL
error
Cure:
Use
the DEBUG command to locate the programming error. Refer to the
HP 70000
Modular
Spectrum
Analyzer Programming Manual
.
2023
Illegal Cal signal
Cause: Calibration signal did not match.
2024
Illegal HP-MSIB comm
Illegal HP-MSIB communication
Cause: An HP-MSIB protocol violation has occurred.
Cure: Document all steps leading up to the error condition and contact your HP representative.
2025
System error (slave)
Cause: HP-MSIB communications with a slave module has created this error
.
Cure:
Record the error number and the hexadecimal code.
Record all events that led up to the occurrence of the error message.
Record the HP model numbers of the modules in the system.
Record the rmware version.