Operating
Errors
2000
{
2999
2047
Preselector disabled
Cause: The preselector was bypassed while the peak command was executed.
Cure: Select the preselector path before executing the peak command.
2048
Userdef protected
Cause: User-dened function or
4
USER
5
key is protected.
Cure: Use the DEBUG command to locate the programming error. Refer to the DEBUG
command description in the
HP 70000 Modular Spectrum Analyzer Programming Manual
.
2049
Battery
failed
Cause: The battery backed up RAM was not valid.
Cure: Return to Hewlett-Packard for servicing.
2050
Ampcr/span
too large
Amplitude-correction
data too large
2051
File not found
Cause: A user state was not found in memory.
Cure: Check the state number or name.
2052
File already exists
2053
storage device
2054
write protect
Cause: Mass storage device is write protected.
6-20 Error Messages