Operating
Errors
2000
{
2999
2040
Partial USTATE data
Cause: A recalled user state is not complete.
2041
CAL POWER lvl err
CAL power level error
Cause: The calibration signal level is not sucient for system calibration.
Cure: Increase the signal level.
0
10 dBm is optimum.
2042
Not
stored,
A
0
X->A on
Cure: Use the DEBUG command to locate the programming error. Refer to the
HP 70000
Modular Spectrum Analyzer Programming Manual
.
2043
LINEAR not allowed
Cure:
Use
the
DEBUG
command
to
locate
the
programming
error
. Refer to the
HP 70000
Modular
Spectrum
Analyzer
Programming
Manual
.
2044
Not
stored:
open
1st
Cause:
The
data
for
the
short
is
entered before the data for the open when normalizing for
swept
response
.
Cure: Store the data for the open rst. Refer to the STORREF command.
2045
HP-IB bus error
Cause: HP-IB protocol violation.
Cure: V
erify cables and computer for proper operation.
2046
No DAC in this band
Cause: There is no preselector in the current band.
Cure: Verify the start and stop frequencies and see the manual for operation ranges.
Error Messages 6-19