Operating
Errors
2000
{
2999
2033
Single band only
Cause: The SIGID command requires that the trace be in a single microwave band.
Cure: Adjust the start or the stop frequency so that the trace does not span more than a single
band.
2034
Test switch on
Cause: The module test switch is in the \test" position. All RAM is erased at each power cycle.
Cure: Set the module test switch to the \normal" position.
2035
Illegal
operation
2036
HP-IB
multiple
cntlr
Cause:
A
DLP
used an OUTPUT command or the spectrum analyzer attempted to access the
disk when another device had control of the HP-IB.
Cure: Remove the other device from the HP-IB.
2037
No
instr
resp
Cause:
No
HP-IB
instrument
response from an OUTPUT command or mass storage (such as
SA
VE
or
RECALL) to an HP-IB disk.
Cure:
V
erify
that the HP-IB address is correct.
2038
Span
4
trace too large
Cause: There are too few data points for the span requested.
Cure: Reduce the span or increase the number of trace points.
2039
User stack overflow
Cause: There is not enough RAM space remaining for the operation requested. Often nested
DLPs are calling too deeply.For example, DLP \A" calls DLP \B" which calls DLP \C", and so
forth.
Cure: DLP program should be redesigned using fewer nesting levels.
6-18 Error Messages