EasyManua.ls Logo

HP 8903B - Page 166

HP 8903B
240 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 8903B
Gain Special
0
1.11
4 1.12
8
1.13
12 1.14
(dB) Function
TROUBLESHOOTING---SERVICE
SHEET
2
Amplitude
Limits
Level
(TTL)
at
U11
Pins
(div PPI
8
9
16
1
1.9
to
2.1
L H
H H
3.0
to
3.3
H
L
H H
4.7
to
5.3
H
H
L
H
7.6
to
8.4
H
H
H
L
Service
TROUBLESHOOTING
General
Procedures for checking the
Input
Amplifier Assembly are given below. The circuits
or
points
to
check
are
marked
on
the schematic diagram by
a
hexagon with
a
check mark and a number inside, for example,
a.
In
addition, any points outside
the
labeled circuit area that must be checked are also identified. Fixed signals are
shown
on
the schematic
also
inside
a
hexagon, for example,
<+1.9
TO
+2.1
VDC
>.
Extend the board assembly
where necessary
to
make measurements. These procediires assume that the soiirce is working properly.
Equipment
Oscilloscope
.
.
.
.
. . . . .
.
. .
.
. .
.
. . . . .
.
.
.
.
.
.
.
.
.
H P 1740A
a
Programmable Gain Amplifier Check
1.
On
the Audio Analyzer.
key
in
41.0
SPCL to initialize the instrument. Set the INPUT and OUTPUT switches
to
ground. Key in AMPTD
1
V. Key
in
1.11
SPCL
to set the input gain to
0
dB. Connect the HIGH OUTPUT
to
the HIGH INPUT.
2. Connect
a
high-impedance. ac coupled oscilloscope
to
A2TP5 (SING END). The waveform should be 2.7
to
2.9 Vpp. (The period of the signal shoiild be approximately
1
nis.)
Hint: If the signal is faulty, see Service Sheet
I
and check the
Differential-to-Single-Ended
Amplifier.
3.
Use the oscilloscope’s vertical gain controls
to
adjust the waveform amplitude for 2 divisions peak-to-peak.
4. Connect the oscilloscope to A2TP4 (PGM AMP). On the Audio Analyzer, key in the Special Functions listed
in Table 8B-16. For each setting the peak-to-peak amplitude should be within the limits indicated. If the signal
level is incorrect, also check the control signals indicated. (Checking U
I
1
will require unplugging the leftmost
plug-on filter A2AI or probing from the circuit side of A2.)
a
Weighting Bandpass and High-Pass Filters Check
1. On the Audio Analyzer,
key
in 41.0 SPCL to initialize the instrument. Set the INPUT and OUTPUT switches
to
ground. Key in AMPTD 1
V.
Key in 1.1 1 SPCL
to
set the input gain
to
0
dE. Connect the HIGH OUTPUT
to the HIGH INPUT.
2. Connect
a
high-impedance, ac coupled oscilloscope to A2TP4
(PGM
AMP). The waveform should be 2.7
to
2.9 Vpp. (The period of the signal should be approximately
I
ms.)
Hint: If the signal is faulty, check the Programmable Gain Amplifier.
3.
Use the oscilloscope’s vertical gain controls
to
adjust the waveform amplitude for 2 divisions peak-to-peak.
4.
Connect the oscilloscope to A2TP3 (OUT).
5.
Set the source frequency
as
listed in Table 8B-17.
For
each setting, press the appropriate filter
key.
The
peak-to-peak amplitude should be within the limits indicated.
8B-2
1

Related product manuals