208
Equivalent Front-Panel Key Commands
Syntax
:INCRement:MULTiply
Example
OUTPUT 714;"RFG:FREQ:INCR:MULT"
If the RF Generator’s frequency increment is 1 MHz, this command
increases increment value from 1 MHz to 10 MHz.
INCR÷10
The INCR÷10 Data Function reduces the increment setting by a factor of 10 (new
increment setting = current increment setting ÷ 10).
NOTE: Integer-only fields (such as Intensity and Print Adrs) have a fixed
increment of 1, and cannot be changed.
Syntax
:INCRement:DIVide
Example
OUTPUT 714;"RFG:FREQ:INCR:DIV"
If the RF Generator’s frequency increment is 10 MHz, this command
reduces the increment value from 10 MHz to 1 MHz.
Increment Up/Down (Arrow Keys)
The Increment Up/Down (Arrow Keys) Data Functions change the field’s setting
by one increment value (up or down). The increment value is determined by the
INCR SET (:INCRement) Data Function.
Syntax
:INCRement <UP or DOWN>
Example
OUTPUT 714;"RFG:FREQ:INCR UP"
This increases the RF Generator’s frequency by one increment value.