230
Common Command Descriptions
*RST (Reset)
The *RST command resets the Test Set. When the *RST command is received the
majority of fields in the Test Set are “restored” to a default value, some fields are
“maintained” at their current state and some are “initialized” to a known state.
Refer to "Instrument Initialization" on page 321 for further information. Other
operational characteristics are also affected by the *RST command as follows:
All pending operations are aborted.
The Test Sets display screen is in the UNLOCKED state.
Measurement triggering is set to TRIG:MODE:SETT FULL;RETR REP.
Any previously received Operation Complete command (*OPC) is cleared.
Any previously received Operation Complete query command (*OPC?) is cleared.
The power-up self-test diagnostics are not performed.
The contents of the SAVE/RECALL registers are not affected.
Calibration data is not affected.
The HP-IB interface is not reset (any pending Service Request is not cleared).
All Enable registers are unaffected: Service Request, Standard Event, Communicate,
Hardware #1, Hardware #2, Operation, Calibration, and Questionable Data/Signal.
All Negative Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operational, Calibration, and Questionable Data/Signal.
All Positive Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operational, Calibration, and Questionable Data/Signal.
The contents of the RAM memory are unaffected.
The contents of the Output Queue are unaffected.
The contents of the Error Queue are unaffected.