EasyManua.ls Logo

HP 8920B - Page 278

HP 8920B
682 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
278
Chapter 5, Advanced Operations
Status Reporting
The following sections show the syntax and give programming examples (using
the HP BASIC programming language) for the Common Commands used to
access the Standard Event Status Register Group’s registers.
Reading the Event Register
Syntax
*ESR?
24 Query
Error
1 = An error has occurred while trying to read the Test Set’s Output
Queue. The following events cause a query error:
a. An attempt is being made to read data from the Output Queue when
no data is present or pending.
b. Data in the Output Queue has been lost. An example of this would
be Output Queue overflow.
12 Request
Control
1 = The Test Set is requesting permission to become the Active
Controller on the HP-IB bus.
0 1 Operation
Complete
1 = The Test Set has completed all selected pending operations and is
ready to accept new commands. This bit is only generated in response
to the *OPC IEEE 488.2 Common Command.
Table 18 Standard Event Status Register Bit Assignments (Continued)
Bit
Number
Binary
Weighting
Condition Comment

Table of Contents

Other manuals for HP 8920B

Related product manuals