Chapter 5, Advanced Operations
Instrument Initialization
The Power-On Reset condition in the Test Set was specifically designed to
configure the instruments for manual testing of an FM radio. The Power-On Reset
default display screen is the RX TEST screen. Other operational characteristics
are also affected by the Power-On Reset as follows:
The Power-up self-test diagnostics are performed.
The Contents of the SAVE/RECALL registers are not affected.
All pending operations are aborted.
Measurement triggering is set to TRIG:MODE:SETT FULL;RETR REP
All Enable registers are cleared: Service Request, Standard Event, Communicate,
Hardware #1, Hardware #2, Operation, Calibration, Call Processing and Questionable
Data/Signal.
All Negative Transition Filter registers are initialized to all zeros: Communicate,
Hardware #1, Hardware #2, Operation, Calibration, Call Processing and Questionable
Data/Signal.
All Positive Transition Filter registers are initialized to all ones: Communicate,
Hardware #1, Hardware #2, Operation, Calibration, Call Processing and Questionable
Data/Signal.
Any previously received Operation Complete command (*OPC) is cleared.
Any previously received Operation Complete query command (*OPC?) is cleared.
Calibration data is not affected.
The HP-IB interface is reset (any pending Service Request is cleared.)
The contents of the RAM memory are unaffected.
The Test Sets display screen is in the UNLOCKED state.