Chapter 8, Programming The Call Processing Subsystem
Programming The CALL CONFIGURE Screen
Sat Tol
This field is used to select the Sat tolerance. When set to Narrow the Sat limit is
one of the three valid Sat tones ± 10Hz. When set to Wide the Sat limit is defined
at 0 to 7000 Hz. When operating in noisy environments the narrow Sat limit may
not allow for correct measurements and the ability to alter the limit to encompass
a wider range provides for testing flexibility.
The :STOLerance command is used to control this field.
The query form (that is, :STOLerance?) can be used to query the current field
value.
Syntax
:STOLerance <‘><NARROW/WIDE><’>
:STOLerance?
Example
OUTPUT 714;"CALLP:AMPS:STOL ‘NARROW’"
OUTPUT 714;"CALLP:AMPS:STOL?"
ENTER 714;Stol_set$
TX Pwr Zero
The TX Pwr Zero function establishes a 0.0000 W reference for measuring RF
power at the RF IN/OUT port. The TX Pwr Zero field is imported from the RF
ANALYZER
screen and is programmed exactly as it is on its home screen. See "RF
Analyzer" on page 177
for programming command syntax.
RF power must not be applied while zeroing the power meter.