659
Error 150 String data error
This error, as well as errors 151 through 158, are generated when parsing a string
element.
Error 151 Invalid string data
A string data element was expected, but was invalid for some reason (see IEEE 488.2,
7.7.5.2). For example, an END message was received before the terminal quote character.
Error 152 Parity error
Parity error
Error 158 String data not allowed
A string data element was encountered but was not allowed by the device at this point in
parsing.
Error 160 Block data error
This error, as well as errors 161 through 168, are generated when parsing a block data
element.
Error 161 Invalid block data
A block data element was expected, but was invalid for some reason (see IEEE 488.2
7.7.6.2). For example, an END message was received before the length was satisfied.
Error 168 Block data not allowed
A legal block data element was encountered but was not allowed by the device at this
point in parsing.