Error −202 Settings lost due to rtl
Indicates that a setting associated with a hard local control (see IEEE 488.2, 5.6.1.5) was
lost when the device changed to LOCS from REMS or to LWLS from RWLS.
Error −210 Trigger error
Error −211 Trigger ignored
Indicates that a GET, *TRG, or triggering signal was received and recognized by the
device but was ignored because of device timing considerations. For example, the device
was not ready to respond.
Error −212 Arm ignored
Indicates that an arming signal was received and recognized by the device but was
ignored.
Error −213 Init ignored
Indicates that a request for a measurement initiation was ignored as another measurement
was already in progress.
Error −214 Trigger deadlock
Indicates that the trigger source for the initiation of a measurement is set to GET and
subsequent measurement query is received. The measurement cannot be started until a
GET is received, but the GET would cause an INTERRUPTED error.
Error −215 Arm deadlock
Indicates that the arm source for the initiation of a measurement is set to GET and
subsequent measurement query is received. The measurement cannot be started until a
GET is received, but the GET would cause an INTERRUPTED error.
Error −220 Parameter error
Indicates that a program data element related error occurred.